CO2 Incubator Detector
Gases Measured
CO2 By custom dual wavelength infra-red with reference channel
O2 (Optional) By internal electrochemical cell
Oxygen cell lifetime Approximately 3 years in air
Range
CO2 0-20%
O2 0-100%
Measurement Accuracy
CO2 ± 1% of range after calibration
O2 ± 1% of range after calibration
Response time T90
CO2 ≤ 20 seconds
O2 ≤ 60 seconds
Temperature (optional) x 2 using optional probes 0°C to +50°C
Temperature accuracy, typical ± 0.1°C from 32 to 44°C, ± 0.2°C over the rest of the range
Barometric pressure 800 - 1200 mbar
RH measurement (optional) RH Probe 0 - 100% RH non condensing
RH accuracy ± 1.5% RH across the range
Visual and audible alarm User selectable CO2 and O2 alarm levels
Communications USB type B mini-connector, HID device class
Data storage 1000 reading sets + 270 events
Operating temperature 0°C to 50°C
Relative humidity 0 - 95% non condensing (RH Probe 0 - 100% non condensing)
Barometric pressure ± 500mbar from calibration pressure
IP rating IP40
Flow 100cc/ min typically
Battery type Li Ion
Battery life 12 hours (10 hours with pump)
Battery lifetime 600 cycles
Battery charger 5v DC external power supply and internal charging circuit
Charge time 4 hours
Alternative power 5Vdc power supply
Weight 495g
Size L 165mm, W 100mm, D 55mm
Case Material ABS/ Polypropylene with Silicone Rubber Inserts
Keys 17 Resin capped Silicone Rubber keys
Display
Liquid crystal display, 128 x 64 pixel
With RGB LED back-light
Gas sample filter
Built-in gas dryer tube to remove moisture
User replaceable PTFE water trap filter